Recall Update: React Health (doing business as 3B Medical) has updated its voluntary correction of the Luna G3 BPAP System, Model LG3800-30VT. The company's update is dated November 17, 2026, and follows its original customer notice of October 8, 2026. The FDA says 7,533 devices were distributed in the US and Puerto Rico. The FDA posts the announcement as a public service and does not endorse the product or company.
What's Being Recalled?
The recall covers the Luna G3 BPAP (Bilevel Positive Airway Pressure) System, a medical device distributed through durable medical equipment (DME) providers, hospitals and healthcare facilities. No retailers or prices are listed. The affected products are identified by:
- Luna G3 BPAP System, Model LG3800-30VT, UDI-DI 06948538365731, serial numbers listed in Appendix A of the FDA notice
The full list of affected lot codes is on the agency page linked below.
Why Was This Product Recalled?
A data storage and transmission error occurs when the "Targeted VT" feature is enabled. The device does not record the actual pressure delivered to the patient on its SD card, and does not send it to the React Health Connect cloud system. Users who have not enabled Targeted VT will not experience the issue. The FDA notice says a clinician reviewing the inaccurate history could adjust settings based on it, which could result in delayed therapy until the setting is corrected. React Health has received 15 complaints, including 2 serious adverse reports, tied to this issue.
What Should You Do?
Do not stop therapy. The company says to contact your health care provider or DME immediately to assess and update your prescribed settings. Do not change settings or discontinue therapy on your own, and keep using the device as prescribed until your provider advises you. Complete the Customer Recall Notification Response Form from the October 8 notice and email it within 7 days to 3B-Recall@reacthealth.com to get the firmware correction (version G3-2.00.80). If you got your device through a DME, ask them to install it. Customer support is 863-226-6285, Monday through Friday, 8:00 a.m. to 5:00 p.m. Eastern. Report problems to FDA MedWatch at 1-800-332-1088.